Monthly Archives: February 2017

New Confirmation Setting in TraceFinder 4.0 and above

TraceFinder has always had a plethora of criteria that is available to assist in the review of data quality. Below is a picture of the Chromatographic Suitability panel that can be set for each peak individually or as a global setting. This flagging criteria has existed since Xcalibur’s creation in 1994. However, unlike Xcalibur, TraceFinder will report out the actual measured value of each set criteria versus the simple PASS or FAIL. These values can be visualized in the data review summary flags or utilized in creating reports.

However, on each type of confirming criteria (Ion Ratios, Isotope Patterns, Library Searches, etc..) there exist the ability to turn them on or off for the method. Added in the TF 4.0 release is the ability to enable the Required option. Setting this option allows the users to control for peaks that only pass the set criteria. All peaks that have the Required option turned on for one or many confirming criteria and have a failed value for the Required option will not be intergrated. Thus, it will be as if the peak was not found in the sample.

This is an optional feature and is turned off by default when the compounds/peaks are created in the Master Method Template.

capture

Tip for Optimizing the Speed of Processing Data for TraceFinder 4.0 and later

TraceFinder 4.0 introduced the ability to perform target analysis on peptide sequences. Given that these chemical analysis methods can be very long (i.e. 3hours for a single sample) and have drastically shifting retention times between samples, TF 4.0 introduced the Extraction Time Window as a peak detection setting.

This new setting allows the user to set a window around where the peak might elute. The default setting is 3 minutes, which means +/- 1.5 minutes from the expected retention time.

This setting may be a bit wide for stable chromatography. A typical setting for the Extraction Time Window is 1 minute or +/- 30 seconds from the expected retention time. This setting can be changed in the Master Method Template or the Local Method.  This new setting can then be pushed to all the compounds with two mouse clicks or can be optimized for each target peak, clicking the Three Lined Button in the right coner of the settings page and selecting “Apply current extraction window settings to all peaks in the method

Since the amount of data being extracted for the targeted peak is reduced, the amount of time per peak analysis is reduced. This setting is separate from the Peak Detection and View Width Window under the Retention Times Tab.

capture

Peak Extraction Window

Poll for Windows 10 usage

Tips on Optimizing Speed for TraceFinder4.0 and later

The first tip for speed optimization can be used on any application. On many systems, the power setting is set for Balanced, especially for laptop users as this conserves battery power. However, graphic intense applictions and those applications that require your processor to be run optimally, need the CPU to be fully powered. Otherwise, the CPU and RAM will not perform at their specified speeds.

Under the Control Panel for Windows 7, select the High Performance Option.

capture

High Performance Power Option

 

Second, Microsoft has put out a new version of the .NET Framework, which drives many of the underlying functions in Windows and applications written for Windows 7. Most people have no idea that this exist.

The new version .NET 4.6.2 fixes many issues Windows was experiencing with .NET 4.5.1, which TraceFinder 3.3 and above was installed with. Also, if your IT group or new PC has .NET 4.6 or 4.6.1, Microsoft has suggested that users install the new .NET 4.6.2.

Below is a link to download the framework from Microsoft. Please check with your IT group before making changes on a corporate PC.

.NET Framework Offline Installer

You can check your version of .NET Framework in the Control Panel/Add Remove Programs/Programs and Features

capture

Add/Remove Programs .NET Version